Science and industry move at a dizzying pace—often times in directions that are unexpected or surprising based on the need of the moment. For the pharmaceutical industry in particular, a Raman Chemical Imaging platform that can adapt to constantly changing materials characterization needs is required for the development and analysis of a virtually unlimited number of compounds having widely varying molecular compositions. FALCON II satisfies this need.
- Par-focal among objective selections
- Par-centricity ± 10µm
- Passive environmental vibration damping
- Optical resolution near diffraction limited to 350 nm.
- Spectral resolution 9 cm -1 FWHM based on Raleigh criteria,
0.5 cm-1 peak shift discrimination.
- Measured Raman range from 244 cm-1 to 5,464 cm-1shift
from laser line.
- High sensitivity deep cooled back-illuminated CCD –70°C
- Optional EMCCD
- Maximum optical magnification 400X
- Excitation source solid-state single frequency, single mode
532 ± 0.1nm 2-watt laser
- High quality color video
- Integration to 8.5 sec. allows use in low light
- DSP control for precise, flexible imaging
- Optical resolution near diffraction limited to 350 nm.
- Fluorescent spectral range 400- 720 nm
- Spectral bandpass 7 nm
- High sensitivity deep cooled CCD –70°C
- Maximum optical magnification 400X
- Excitation wavelength 365 nm, derived from Hg arc lamp and Excitation filter
- Dispersive spectral resolution 3 cm-1 FWHM based on Raleigh criteria
- Measured Raman range from 244 cm-1 to 4,454 cm-1 shift from laser line.
- Automated three grating turret Automated slit width
- High sensitivity deep cooled back-illuminated CCD –70°C
- Excitation source same as Raman Imaging
- Patented Deut Vision Technology© enables simultaneous Raman imaging and Raman dispersive measurements
|
|
- Optical resolution 760 nm (@ λ=1μm)
- NIR spectral range 900 –1700nm
- Spectral resolution 7nm o InGaAs FPA CCD camera
- Maximum optical magnification 200X
- Excitation source – Quartz tungsten Halogen, 100W
- Basic dimensions – 92cm deep x 69cm wide x 191cm tall
- Weight ~295 kg (Max.)
- Sample stage travel X and Y = 10cm x 10cm
- Sample X and Y positioning resolution – 0.05 μm
- Sample Z resolution – 0.1 μm
- Sample formats - accepts 25mm x 75mm, 50mm x75mm microscope slides, and standard well plate
|